Skip to content

Transsonic TI-H Manual & Operating Instructions

Operating guidance for the Techspan Transsonic TI-H range of ultrasonic cleaning units —
TI-H 5, 10, 15, 20, 25, 55, 80, 115 and 160.
Taken from the operating instructions supplied
with the unit by Techspan.

Download: Transsonic TI-H Operating
Instructions (PDF)
— full manual including safety warnings, cleaning media limits and maintenance.

Single-frequency and multi-frequency units

TI-H units were supplied in two variants, and which one you have determines what the controls do:

  • SF1 — single frequency, 35 kHz. No frequency selector, no mode selector.
  • MF2 — multi-frequency, 25 / 45 kHz. Frequency and function knobs are fitted.

Check the nameplate if you are not sure. The degas and Sweep functions described below apply to
multi-frequency units only.

Frequency selection (multi-frequency units)

  • 25 kHz — removing coarse and tenacious contamination, and pre-cleaning robust
    surfaces.
  • 45 kHz — fine cleaning and removing contamination from sensitive surfaces.

The frequency can be changed while ultrasound is running.

Standard and Sweep modes (multi-frequency units)

Set at the function knob, and switchable during operation:

  • Standard — use when the basket is connected to a manual or automatic oscillation
    device.
  • Sweep — continuously shifts the sound-pressure maxima for a more homogeneous
    sound field. Use this when the basket is not on an oscillation device.

Technical specifications

ModelMax. tank volume (L)Tank service volume (L)Tank inner W×D×H (mm)Unit outer W×D×H (mm)Basket inner W×D×H (mm)Weight (kg)
TI-H 54.73.5240×130×150340×300×370200×100×9010.5
TI-H 1010.88.6300×240×150400×415×370260×210×8514.8
TI-H 1514.412.2300×240×200400×415×420260×210×13016
TI-H 2019.816.8330×300×200430×470×420290×270×13020
TI-H 2525.519.8330×300×230615×690×570290×270×14032
TI-H 5557.845500×300×350620×850×570255×460×21035
TI-H 8082.567.5500×300×500620×850×720255×460×29060
TI-H 115115.590600×500×350840×1000×570450×545×205100
TI-H 160160135600×500×500840×1000×720440×540×350140
ModelPower consumption total (W)Ultrasonic power effective (W)Ultrasonic peak max. (W)Heating power (W)
TI-H 5500100400400
TI-H 101000200800800
TI-H 1512002008001000
TI-H 20165025010001400
TI-H 25130030012001000

Mains connection 100–120 V or 200–240 V, single phase, with earthing contact.
Ultrasonic frequency SF1 35 kHz or MF2 25/45 kHz depending on variant.

Safety and maintenance

  • Solvent-containing and aqueous cleaners both have limits on use — check section 8 of the manual
    before introducing any new chemistry.
  • On TI-H 25 and larger, the safety switch-off must be checked periodically.
  • The transducer tank is a wearing part; service life depends on the media used and how the unit is
    run.
  • Repairs must be carried out by authorised personnel only.

Support and replacement

The Transsonic TI-H is a legacy range. We continue to support units in the field in New Zealand
and can advise on servicing, consumables, baskets and a current-model replacement sized to your
existing tank. Call 0800 603 603, email
info@elma-ultrasonic.co.nz, or see
service & support.

For a like-for-like size in the current range, look at the
Elmasonic X-tra series for bench units and the floor-standing
X-tra ST series for the larger TI-H tank sizes.

Before replacing a unit that seems weak, confirm it with a
SonoCheck activity test vial.

Download the full Transsonic TI-H Operating Instructions (PDF) · All Elma manuals & downloadsOpens in a new tab

Not sure which machine?

Tell us what you are cleaning, how much of it and how often. We will specify the tank, frequency and chemistry.

Request a quote
Shopping cart0