Skip to content

Elmasonic EASY Manual & Operating Instructions

This page covers the Elmasonic EASY range — EASY 10, 10H, 20, 20H, 30, 30H, 40H, 60H,
100H, 120H, 180H and 300H.
The detail below is taken from the Elma operating instructions
supplied with the unit in New Zealand by Techspan.

Download: Elmasonic EASY Operating
Instructions (PDF)
— the full manual, including installation, fault messages and maintenance.

How the controls work

EASY units are deliberately simple: a cleaning-duration rotary knob and a Start/Stop button.
There is no frequency selector and no Sweep or Pulse mode — every EASY unit runs at a single
37 kHz.

  • Off — rotary knob turned to OFF, all LEDs off.
  • On — knob turned to ON, or a cleaning duration of 1–30 minutes selected. The LED
    for the set ultrasound mode lights.
  • Cleaning — press Start/Stop. The cleaning-duration LED flashes while ultrasound
    runs. Press Start/Stop again at any point to end the cycle.
  • Complete — a short audible signal sounds at the end of the set duration and the
    cleaning-duration LED returns to steady.
  • Safety switch-off — the unit shuts down automatically after 8 hours of inactivity
    or after a brief power interruption. Press any button or turn the knob to wake it.

Before your first cycle

Degas every fresh bath. New liquid holds dissolved air that absorbs ultrasonic energy and suppresses
cavitation, so cleaning is poor until the gas is driven out. Run the unit for roughly 10 minutes with
the liquid in and the basket out, then load your work.

Installation requirements

The EASY range is designed for commercial premises such as laboratories and medical practices. The
manual is specific about siting:

  • Well-ventilated location.
  • A sturdy, horizontal, non-slip, moisture-resistant and non-flammable work surface.
  • Enough clearance above the unit to lift the cover and the items out.
  • Space beside the unit to set the basket and inserts down.
  • A splash-proof earthed socket near the unit — the cable is about 1.5 m.
  • The supply must be protected by a residual-current device (RCD).

Technical specifications

EASY 10, 10H, 20, 20H

 EASY 10 / 10HEASY 20 / 20H
External W×D×H with cover (mm)235×130×200215×190×230
Interior W×D below fill mark (mm)185×80140×125
Basket interior W×D×H (mm)155×70×35110×105×50
Total volume (L)0.91.6
Recommended operating capacity (L)0.71.2
Weight with cover (kg)2.02.1
Ultrasonic power effective (W)3035
Ultrasonic peak max. (W)240280
Heating power (W)– / 60– / 120

EASY 30, 30H, 40H, 60H

 EASY 30 / 30HEASY 40HEASY 60H
External W×D×H with cover (mm)305×190×230305×190×280370×225×280
Basket interior W×D×H (mm)195×105×50195×105×95255×105×95
Max. basket loading (kg)1.03.05.0
Total volume (L)2.73.95.9
Recommended operating capacity (L)1.62.94.3
Weight with cover (kg)3.34.05.1
Ultrasonic power effective (W)80120150
Heating power (W)– / 200200400

EASY 100H, 120H, 180H, 300H

 EASY 100HEASY 120HEASY 180HEASY 300H
External W×D×H with cover (mm)380×315×280380×315×330405×375×330575×375×330
Basket interior W×D×H (mm)250×195×90250×195×135280×255×135450×255×135
Max. basket loading (kg)6.07.08.010.0
Total volume (L)9.512.917.827.5
Recommended operating capacity (L)6.810.014.222.0
Weight with cover (kg)5.97.58.511.0
Total power consumption (W)550100010001500
Ultrasonic power effective (W)150200200300

Ultrasonic frequency across the EASY range is 37 kHz (tolerance −2.5 / +5.5 kHz).
Effective ultrasonic power figures carry a ±10% tolerance. “H” models are heated.

Maintenance

Keep the enclosure and tank clean, disinfect where your protocol requires it, and on sizes 120 and
up check that the blower is clear and running. The mains cable is replaceable. Anything beyond routine
operation and cleaning must be done by a qualified, authorised technician.

If cleaning performance drops off, confirm the bath is actually cavitating before you assume the
unit is faulty — a SonoCheck test vial gives you a pass or fail in one
cycle.

Frequently asked questions

What frequency is the Elmasonic EASY?

37 kHz on every model in the range. Unlike the P and Select ranges, EASY units do not offer a
switchable 80 kHz setting.

How long can I set the timer for?

1 to 30 minutes, or continuous by turning the knob to ON. A short beep marks the end of a timed cycle.

My unit turned itself off — is it faulty?

Probably not. EASY units switch off automatically after 8 hours of inactivity or after a brief
power interruption. Press a button or turn the rotary knob to wake the unit.

Which cleaning solution should I use?

It depends on the contamination and the substrate, not the machine. We stock the
Elma TEC CLEAN range in New Zealand — tell us
what you are cleaning
and we will match a concentrate and dilution.

Buy or replace an EASY unit

We stock the full Elmasonic EASY series in New Zealand:
EASY 30H,
EASY 40H,
EASY 60H,
EASY 100H,
EASY 120H,
EASY 180H and
EASY 300H.
Call 0800 603 603 or email info@elma-ultrasonic.co.nz.